| Herkunftsort: | Niederlande |
|---|---|
| Markenname: | YOKOGAWA |
| Zertifizierung: | CE |
| Modellnummer: | SC8SG-R61-F-313-Y1*A |
| Dokumentieren: | SC210G SC8SG SC4AJ GS12D08G...10.pdf |
| Min Bestellmenge: | 1 Stk |
| Preis: | verhandelbar |
| Verpackung Informationen: | Originalverpackung |
| Lieferzeit: | 1 Woche |
| Zahlungsbedingungen: | T/T |
| Versorgungsmaterial-Fähigkeit: | 100 Stück pro Monat |
| Technischer Name: | Leitfähigkeits-Detektor | Messbereich: | Hohe Reichweite; Zellkonstante: 10 cm-1 |
|---|---|---|---|
| Elektrodenkonfiguration: | 4-Elektroden-System | Durchflusstyp: | 1/2NPT Innengewinde; Kammermaterial: PP |
| Kabellänge: | 5.5 m | Herkunft: | JAPAN |
| Hervorheben: | YOKOGAWA conductivity sensor with warranty,YOKOGAWA transmitter conductivity sensor,SC8SG-R61-F conductivity sensor |
||
The SC8SG conductivity sensor is widely used in applications such as liquid conductivity measurement for manufacturing processes and pure water resistivity measurement in the semiconductor, food, pharmaceutical, and power industries. A wide variety of mountings are available including screw-in type, flow-through type with screw joint, and flow-through type with flange.
Specifications
| Object of measurement | Conductivity of liquids or solutions |
| Measuring Principle | 2-electrode or 4-electrode system |
| Cell Constants | 0.01 cm-1 or 10 cm-1 (for 2-electrode system) 10 cm-1 (for 4-electrode system) |
| Measuring Ranges | 0 - 0.5 μS/cm to 0 - 100 μS/cm for a cell constant of 0.01 cm-1 0 - 1 mS/cm to 0 - 1000 mS/cm for a cell constant of 10 cm-1 |
| Temperature Range | 0 to 100 deg C (130 deg C maximum only for 0.01 cm-1 cell constant detectors, excluding those with polypropylene chambers) |
| Pressure | 1000 KPa max. (500 kPa maximum for detectors with polypropylene chambers) |
| Flow rate of Sample Solution | No particular limitation applies, although a value of less than 20 l/min is recommended for flow-through detectors. |
| RTD for Temperature Compensation | Pt1000 (built into the sensor) |
| Process Connection | Screw-in or flow-through |
![]()
![]()
![]()